Parameter measured

  • Temperature control range: 0˚C to 90˚C +/- 0.1˚C.
  • Condensation control: Purge using dry air.
  • Standard laser: 4mW, 633nm.
  • Correlator: 25ns to 8000s, max 4000 channels.

Size

  • Absolute sensitivity (Toluene kcps): 150.
  • Range (Maximum diameter): 0.3nm - 10 microns.
  • Min sample volume: 12µL.
  • Min concentration, protein: 0.1mg/ml 15kDa protein.
  • Min concentration, forward angle: 10mg/ml 66kDa protein.
  • Max concentration: 40% w/v.
  • Measurement angles: 13˚ + 173˚.

Zeta potential

  • Sensitivity: 10mg/mL 66kDa protein.
  • Zeta potential range: > +/- 500mV.
  • Mobility range: > +/- 20µ.cm/V.s.
  • Maximum sample concentration: 40% w/v.
  • Maximum sample volume (using diffusion barrier): 12µL.
  • Maximum sample conductivity: 200mS/cm.
  • Signal processing: M3-PALS.

Molecular weight

  • Molecular weight range (estimated from DLS): <1000Da - 2x10^7Da.
  • Molecular weight range (Debye plot): <1000Da - 2x10^7Da.